The Cardiac Science 5350
The Cardiac Science 5350 vital signs monitor
NEW! The Cardiac Science 5350 vital signs monitor gives you all the functionality of the Cardiac Science 5300 – plus alarm settings and intervals for continuous monitoring.Choose the 5350 to view and monitor noninvasive blood pressure, pulse rate, temperature, and SpO2 efficiently and comfortably with a big screen and bright, color-coded digits.
How the 5350 stacks up versus the competition
- The screen is nearly 3x larger and noticeably brighter than a major competitor’s.1 With the large screen , 45-degree slanted screen, and bright, color-coded digits, you'll be able to focus on the patient and the monitor.
- The wide base makes it difficult to tip over, so costly temperature probe damage may be a thing of the past.
- The easy-to-clean monitor withstands routine fluid contact to help contain the spread of infection.
- The Smart Inflation™ technology customizes cuff inflation for each patient to the minimum pressure required for an effective test. Paired with dynamic linear deflation (which deflates the cuff based on the patient's pulse rate), the test makes for more comfortable, happier patients.
- The optional built-in printer lets you easily document the test.
For continuous monitoring
Perfect for continuous monitoring, the Cardiac Science 5300 is equipped with an audible alarm to sound should a patient's measurement fall outside the pre-programmed range. (There is also a visual indicator.)A wide variety of timer intervals can automatically take timed blood pressure measurements. Users can select interval measurements ranging from continuous to every 180 minutes.
Benefits
- Better visibility. The 5350 is designed with a screen nearly 3x larger and noticeably brighter than a major competitor’s to enhance staff efficiency and accuracy.1
- Durability. Like all Cardiac Science devices, these vital signs monitors are built to last, even in a busy hospital environment.
- Patient comfort. Smart Inflation™ technology prevents the blood pressure cuff from overinflating and, with dynamic linear deflation, the cuff deflates based on the patient’s pulse rate.
- Ease of use. The 5350 withstands routine fluid contact so cleanup is fast and easy.
- Connectivity. For EMR integration, an optional connectivity kit enables you to download data in HL7 format or network service to a laptop.
- Easy documentation. The optional built-in printer allows you to print out results immediately or later. Select among three report styles.
1 Compared to the Welch Allyn® Spot Vital Signs® 420 series monitor, according to the Directions for Use REF 4200-87E, Material No. 706272 Ver. C
|
NONINVASIVE BLOOD PRESURE (NIBP) |
|
Measurement technology: |
Oscillometric method |
Measurement method: |
Linear deflation method |
Measurement time: |
Normal / High speed selectable |
Pressure display range: |
0 mmHg to 299 mmHg |
Pressure display accuracy: |
±3 mmHg |
NIBP Measurement range: |
- |
Adult/pediatric mode - SYS: |
60 mmHg to 250 mmHg |
Adult/pediatric mode - MAP: |
45 mmHg to 235 mmHg |
Adult/pediatric mode - DIA: |
40 mmHg to 200 mmHg |
Adult/pediatric mode - Pulse rate: |
40/min to 200/min |
Neonatal mode - SYS: |
40 mmHg to 120 mmHg |
Neonatal mode - MAP: |
30 mmHg to 100 mmHg |
Neonatal mode - DIA: |
20 mmHg to 90 mmHg |
Neonatal mode - Pulse rate: |
40/min to 240/min |
NIBP accuracy: |
ANSI / AAMI SP-10 |
Hose Connector: |
RECTUS |
Pulse rate accuracy: |
±2% or ±2 beats |
|
NELLCOR SpO2 OPTION (PULSE OXIMETER) |
|
Measurement Method: |
2 wave length pulse wave type (OXIMAX® technology) |
Measurement Range: |
70% to 100% SpO2 |
Measurement Accuracy: |
±2% SpO2 (70% to 100% SpO2 when using SpO2 disposable sensor MAX-A) ±3% SpO2 (70% to 100% SpO2 when using SpO2 reusable sensor DS-100A) |
Pulse Rate Range: |
20 to 250/min |
Pulse Rate Accuracy: |
±3/min |
Display Update: |
Less than 30 sec |
|
MASIMO SpO2 OPTION (PULSE OXIMETER) |
|
Measurement Method: |
2 wave length pulse wave type (Masimo SET® technology) |
Measurement Range: |
1% to 100% SpO2 |
Measurement Accuracy: |
±2 SpO2 (70 to 100% SpO2, Adult/Pediatric, without motion) ±3 SpO2 (70 to 100% SpO2, Neonatal, without motion) |
Pulse Rate Range: |
25 to 240/min |
Pulse Rate Accuracy: |
±3/min |
Display Update: |
Less than 30 sec |
|
ALARIS TEMPERATURE OPTION |
|
Measurement Method: |
Turbo Temp® Electronic Predictive Thermometer |
Measurement Site: |
Sub lingua and Axilla |
Measurement Range: |
35.6°C - 41.1°C (96.1°F - 105.9°F) (Predictive mode) 26.7°C - 41.1°C (80.1°F - 105.9°F) (Monitoring mode) |
Measurement Accuracy: |
±0.1°C (±0.2°F) (Monitoring mode) |
|
COMMUNICATIONS KIT |
|
Type: |
RS-232 Port |
Output: |
HL7 or ASCII |
|
PRINTER OPTION |
|
Type: |
Thermal line head |
Reports: |
Simple list, detail list, or measurement value |
Resolution: |
8 dot/mm |
Paper Type: |
Thermal |
Paper Width: |
58 mm (2.28 in) |
Print Speed: |
25.0 mm/sec |
|
GENERAL |
|
Dimensions (H x D x W): |
- |
Main Unit: |
150 mm x 239 mm x 239 mm (5.9 in x 9.4 in x 9.4 in) |
AC Adaptor: |
75 mm x 47 mm x 150 mm (2.9 in x 1.8 in x 5.9 in) |
Weight, Main Unit: |
3.6 kg (8.1 lb) (including all options but excluding accessories) |
Display Method: |
LED |
Type of protection: |
Type BF - Applied part |
Protection Class: |
Class I |
|
POWER REQUIREMENT |
|
AC Power: |
AC 100 V~240 V, 50/60 Hz |
Battery: |
12 V, 3.2 Ah |
Battery - Type: |
Lead acid |
Battery - Operating Time: |
6 hours when fully charged, with no printing, 1 NIBP measurement every 15 minutes, in power-saving mode, at 25°C (77°F) |
Battery - Charging Time: |
6 hours from fully empty to fully charged |
Battery Saving Mode Function: |
Yes |
|
ENVIRONMENTAL |
|
Operating Temperature: |
0°C to 40°C (32°F to 104°F) |
Storage Temperature: |
-20°C to 60°C (-4°F to 140°F) |
Operating Humidity: |
30% to 85% (non-condensing) |