Skip to content

Contact Us

Burdick Spirocard Spirometer

The Burdick SpiroCard is a single PCMCIA card spirometer that provides diagnostic testing in an easily portable package.

  • SpiroCard inserts into a standard PC-card slot on a laptop or PC
  • Delivers full-function spirometry—performs all measurements performed on conventional devices
  • Spirometry graphs and reports automatically created
  • Tests are clear, accurate, and concise
  • Uses patented, disposable and biodegradable patient spirometry filters
  • Interfaces with Office Medic software for real-time data access, reporting, and archiving
  • Warranty: 3-Year Parts and Labor

Please give me more information about
the Burdick Spirocard Spirometer

My name:
My phone:
My email:
What I'm looking for:
Call 800.679.8703 or
 

SPIROCARD SPECIFICATIONS

Dimensions:

2.09" W x 5.51" D x 1.02" H (53 mm x 140 mm x 26 mm)

Weight:

85 grams (less than 3 ounces)

Certifications:

UL CUL listed

Warranty:

Device 3 years return for exchange; Sensors 90 days

 

OXIMETER SPECIFICATIONS

Ranges:

Percentage of saturation: 0 to 100%; pulse rate: 18 to 300 pulses per minute

Accuracy:

SpO2 (+/- 1 SD) = 70-100% +/- 2% of full scale; Pulse rate: +/- 3% (+/- 1 digit)

Record time:

Up to 24 hours

Parameters:

SpO2; pulse rate

 

SPIROMETRY SPECIFICATIONS

Ranges:

FLOW: +/- 14 liters/second

Volume:

0 – 8 liters BTPS

Accuracy:

Complies with American Thoracic Society (ATS) Standards for Spirometry, 1994 Update

Flow:

+/- 5% of indication or 200 ml/sec, whichever is greater for FEF 25-75, and +/- 10% or 400 ml/sec whichever is greater for PEF

Volume:

+/- 3% or 50 ml, whichever is greater for FVC and FEV1

Parameters measured:

FVC, FEV0.5, FEV1, FEV1/FVC, FEV3, FEV3/FVC, PEFR, FEF25%, FEF50%, FEF75%, FEF25-75%, FIVC, FIV0.5, FIV1, FIV1/FIVC, FIV3, FIV3/FIVC, PIFR, FIF50%, FIF 25-75%, FIF.21.2, FVC/FIVC, PIFR, Ext.Vol., MVV, RR, MTV, SVC

BTPS conditions:

Comply with American Thoracic Society Standards for Spirometry, 1994 Update

Elevation corrections:

0 to 15,000 feet (0 – 4,572 meters)

Calibration:

ATS 3-speed or standard calibration check

Predicted normals:

Crapo/Cherniack/Morris/Knudson 1983 Polgar/Hsu/Roberts/Warwick/ECCS

Tests performed:

FVC, pre/post testing, flow volume loop, MVV, SVC

Measuring time:

FVC, up to 30 seconds

Printed scale:

Flow-Volume: (vertical) .5 cm/L/S; (horizontal) 1 cm/1L Volume-Time: (vertical) 1 cm/iL; (horizontal) 1 cm/second

 

COMPUTER REQUIREMENTS

Office Medic Software:

Desktop or laptop PC: Operating system: Windows® 98/2000/XP Home or Pro. SpirOxCard may not fit in all internal PC slots; Office Medic minimum requirements equal to operating system minimum requirements; Type II PC Card Slot; The SpiroCard may not fit into some desktop computers

Cardiac Science Sensor Pressure Tubes Sensor Pressure Tubes for use with SpiroOxCard and SpiroCard

For use with SpiroOxCard and SpiroCard sensors

Image Unavailable Mouthpieces for SpirOxCard or SpiroCard spirometer

Patented, disposable, and biodegradable plastic patient mouthpieces.